SIU 2024: (CO-01) Optimal Placement of Access Sheath in RIRS
Speaker: Dr. K Nanjappa, India
Important Takeaways
-
Access Sheath Placement Benefits and Risks: Proper placement reduces intrarenal pressures, improves irrigation and visualisation, and prevents scope damage. Misplacement, however, can result in ureteral trauma and instrument damage.
-
Selection Criteria for Access Sheath: Choosing sheath size depends on patient anatomy, scope size, and ureteral calibre, with visual calibration for ureteral size being optimal.
-
Placement Techniques for Safety and Efficiency: Proper technique includes using fluoroscopic guidance to position the sheath approximately 2 cm below the pelviureteric junction (PUJ), with successful placement ideally achieved within 30 seconds.
-
Flexible Access Sheath and High-Flow Areas: For flexible sheaths, both the access sheath and flexible scope should reach the area of interest, creating a “high flow” for optimal stone clearance.
-
Exit Strategy and Reuse Risks: Removing the sheath under direct visualisation is recommended to prevent ureteral trauma. If reuse causes a sheath to stick, cutting the sheath at the meatus and retracting it by pulling the steel wire helps in safe removal.
Key Highlights
Access Sheath Advantages and Placement Risks:
Dr. Nanjappa opened his discussion by highlighting that access sheaths in RIRS serve multiple purposes. They improve intrarenal pressures, enhance irrigation flow, increase scope deflection and manoeuvrability, and add a layer of protection for reusable scopes. However, poorly placed sheaths can cause complications, including ureteral trauma, high procedural costs, and, potentially, scope damage.
Selecting the Access Sheath: The key to successful placement is in selecting the optimal access sheath based on:
-
Patient Factors: Height, age, and body habitus.
-
Scope Size Compatibility: Ensuring sheath diameter aligns with the scope’s external diameter.
-
Ureteral Calibration: Optical calibration techniques are recommended to assess ureteral calibre accurately, which allows for selection of the appropriate sheath size:
-
A 6.5 Fr ureteroscope reaching the PUJ allows use of an 11/13 Fr sheath.
-
An 8.9 Fr ureteroscope allows a 12/14 Fr sheath.
-
For a 4.5/6.5 Fr scope, no access sheath may be suitable, necessitating a change in surgical approach.
Matching Sheath to Scope: Dr. Nanjappa introduced the RESD Ratio (Ratio of External Scope Diameter to Internal Sheath Diameter) for optimised pairing, explaining:
-
Ratios below 0.75 indicate a well-matched sheath.
-
Ratios above 0.85 present a mismatch risk. He further simplified selection by advising a 2-Fr difference between the sheath’s inner diameter and scope’s outer diameter for safety.
Techniques for Safe Placement: Dr. Nanjappa outlined essential tips for optimal sheath placement:
-
Guide Wire Use: Placement over a guidewire is essential, with well-lubricated sheaths to facilitate passage.
-
Avoiding Forceful Insertion: Ensuring smooth insertion without resistance prevents ureteral injury.
-
Placement Depth: Under fluoroscopic guidance, positioning the sheath 2 cm below the PUJ is ideal. Dr. Manoj Monga’s suggestion that placement should not exceed 30 seconds helps in avoiding forced insertions and possible damage.
High-Flow Concept with Flexible Access Sheath
For flexible sheaths, placement strategies diverge from traditional sheaths. Here, positioning both the access sheath and flexible scope in “high-flow” zones maximises irrigation and efficiency. Dr. Nanjappa noted that keeping the access sheath slightly behind the flexible scope may limit effectiveness, so full insertion to the region of interest is recommended.
Enhancing Ureteral Readiness with Pre-Stenting and Alpha Blockers
To increase the ease of sheath insertion, Dr. Nanjappa recommended pre-stenting where feasible, as this enlarges and prepares the ureter. Although the stent size or laterality does not significantly impact outcomes, even short-term pre-stenting can ease sheath placement. Use of alpha-blockers before the procedure may also aid ureteral dilation and improve procedural outcomes.
Exit Strategy and Sheath Reuse Complications
Safe removal of the access sheath is equally important. Dr. Nanjappa suggested always removing the sheath under direct visualisation to avoid inadvertent ureteral injury. Reuse of sheaths, which is common in certain regions, increases the risk of sheaths sticking within the ureter, especially over prolonged durations. In such cases, cutting the sheath at the meatus and retracting it via the embedded steel wire helps release the sheath by collapsing it, avoiding trauma during extraction.
Summary Recommendations:
-
Avoid routine access sheath placement; consider it only when necessary based on scope size and ureteral calibration.
-
Never force sheath insertion and adhere to the 30-second placement guideline.
-
Pre-stenting and alpha-blockers are recommended adjuncts to facilitate safe sheath insertion.
-
Employ a controlled exit strategy with visual guidance, avoiding reuse where possible to mitigate risks.
Société Internationale d'Urologie Congress, 23-26 October 2024, New Delhi, India.